🦴 X-Ray Attenuation
🔊 Ultrasound Reflection
Principles of Medical Imaging
Diagnostic imaging relies on the interaction between radiation (or waves) and human tissue. Different tissues absorb or reflect energy differently, creating contrast in the resulting image.
1. X-Ray Attenuation
As X-rays pass through a medium, their intensity decreases exponentially. This is known as the attenuation law.
Where μ is the linear attenuation coefficient and x is the thickness. Tissues like bone have a high μ compared to soft tissue, appearing lighter on a film.
2. Ultrasound Interpretation
Ultrasound uses sound waves (typically 1–20 MHz). When a wave hits a boundary between two tissues of different Acoustic Impedance (Z), some energy is reflected.
The intensity reflection coefficient (R) is given by:
Deep Dive: Worked Examples
✅ Example 1: Basic Ultrasound Reflection
Calculate the reflection coefficient R between muscle (Z₁ = 1.7 × 10⁶) and bone (Z₂ = 6.4 × 10⁶).
✅ Example 2: X-Ray Half-Value Thickness (HVT)
If the linear attenuation coefficient of a material is 0.5 cm⁻¹, calculate the thickness required to reduce intensity by half.
✅ Example 3: PET Scanner Annihilation
In PET, a positron annihilates with an electron. Calculate the energy of each of the two resulting photons.
✅ Example 4: Acoustic Impedance of Fat
Fat has a density of 920 kg/m³ and the speed of sound in fat is 1450 m/s. Calculate its acoustic impedance.
✅ Example 5: X-Ray Contrast Improvement
Contrast agent Iodine (μ = 25 cm⁻¹) is used. If 2mm of Iodine is present, calculate the ratio of I/I₀.